Ametek Chandler Engineering 1910 Instruction Manual page 15

Pressurized curing chamber
Table of Contents

Advertisement

Heater: When active, the Heaters are enabled. This button and the physical Heater button
must both be active to enable heating.
Home Screen Indicator Controls
Channel Label
Dial Indicator
Set Point Indicator
Auto / Manual Mode
Channel Label: The name given to this channel. This can be changed on the Indicators tab
of the Configuration Screen.
Dial Indicator: A visual representation of the current value relative to the maximum value
for the Channel.
Set Point Indicator: A visual representation of the current set point relative to the maximum
value of the Channel.
Auto / Manual Mode: Displays the current control mode.
: Indicates Manual Mode, the Set Point is in "Percent Output" (0-100% for
Temperature or ±100% for Pressure).
: Indicates Automatic Mode, the Set Point is in Units. When a Schedule is started,
the control is automatically placed in Automatic Mode.
Units: the currently calibrated units of the Channel.
Current Value: a live reading of the input.
Output Indicator: Indicates the current action of the Control Loop. When heating, a flame
icon appears. On the Pressure Indicator, this icon matches the Pump icon when Pressurizing
and the Press Release icon when de-pressurizing.
SECTION 2 – OPERATION
Units
Current Value
Set Point
Output Indicator
Channel Plot
2-3