► Therefore, keeping the following minimum distances to these HF com­
munication devices is recommended:
• Mobile phone GSM 850/GSM 900: 0.99 m
• Mobile phone GSM 1800/GSM 1900/UMTS: 0.7 m
• DECT cordless phones incl. base station: 0.35 m
• WiFi (routers, access points,...): 0.22 m
• Bluetooth devices (third-party products not approved by Ottobock):
0.22 m
NOTICE
Improper product care
Damage to the product due to the use of incorrect cleaning agents.
► Only clean the product with a damp cloth and mild soap (e.g.
453H10=1 Ottobock DermaClean).
► Do not bring the surface of the electrodes into contact with a special
cleaner for Ottobock cosmetic gloves, e.g 640F12.
5 Scope of Delivery and Accessories
Scope of Delivery
1 pc. 13E202=50/13E202=60 suction socket electrode
1 pc. 13E80 adjustment tool
1 pc. 647G334 instructions for use (qualified personnel)
Additional scope of delivery for socket fabrication
1 pc. 13E203 casting template for inner socket
1 pc. 13E204 lamination template for outer socket
2 pc. 503F3 socket screw with Allen head
2 pc. 507S15 lamination plate with bore
The following products are also required:
757M11=X-Change MyoBoy
13E129=* electrode cable
1 pc. 617H46 bonding agent