Installing/mounting
3.3 Safe installation
3.2.1.1

Direct sunlight

NOTICE
Direct sunlight
Damage to device.
The device can overheat or materials become brittle due to UV exposure.
• Protect the device from direct sunlight.
• Make sure that the maximum permissible ambient temperature is not exceeded. Refer to
3.2.2

LDS 6, optics housing CD 6

External stresses and loads
Damage to device by severe external stresses and loads (e.g. thermal expansion or pipe
tension). Process media can be released.
• Prevent severe external stresses and loads from acting on the device.
3.3

Safe installation

Approval
The concept of the Ex approval is that the central unit is not modified, and that special ATEX/
UKEx/IEC Ex-approved cables and bushings are used, and that a special Ex optics housing pair
(CD 6 Ex) is used in the hazardous zone. In addition to this an explosion protection barrier is
added before entry into the hazardous zone. An absolute condition for the approval is that
the equipment is set up according to the system drawing in section Connection diagrams
(Page 44).
The Ex-relevant ingress protection (IP65) comprises the entire area marked with a dashed line
in the following figure.
24
the information in Technical data (Page 59).