Advanced Echo Processing (P810 To P853) - Siemens interranger DPS300 Instructional Manual

Table of Contents

Advertisement

Advanced Echo Processing (P810 to P853)

Note:
The following Echo Processing parameters are intended for use by authorized
Siemens Service personnel or Industrial Instrumentation Technicians familiar with
Siemens ultrasonic echo processing techniques.
Anatomy of an Echo Profile
The relevant parts of an echo profile are listed here. These are visible in either
Dolphin Plus or an oscilloscope.
P810 Scope Displays
Use this feature to monitor the results of Echo Processing parameter alterations.
Connect an oscilloscope to Display Board TP4, TP5, and TP6 (GND, SCOPE and
SYNC).
Observe any combination of the following Scope Displays:
DISPLAY
Echo Profile
TVT Curve
Echo Marker
Echo Lock Window
7ML19981BC01
echo profile
TVT curve
echo marker
83
selected echo
echo lock window
DISPLAY SYMBOL
P
C
n
u
InterRanger DPS300