Commissioning; Connection; Electrical Operating Conditions; Connecting The Electromatic - KaVo ELECTROmatic M Instructions For Use Manual

Hide thumbs Also See for ELECTROmatic M:
Table of Contents

Advertisement

Instructions for use ELECTROmatic M/C and PM/PC
5 Commissioning | 5.1 Connection

5 Commissioning

Note
The ELECTROmatic may be operated exclusively with the
INTRA LUX KL703 LED motor (Mat. no. 1.007.0150), the COMFORTdrive
motorized contra-angle handpiece and the type 4882 power supply.

5.1 Connection

5.1.1 Electrical operating conditions

Property damage due to incorrect pressures.
Property damage to motor or handpiece.
▶ Set the pressures according to the technical specifications.
Property damage due to bad media.
Property damage to motor or handpiece.
▶ Make sure that the compressed air is dry and free of dirt and oil according
to EN ISO 7494-2.
▶ Make sure that the pH of the water is between 7.2 and 7.8.
Note
If the water is hard (above 12 °dH), a water softening device based on an
ion-exchange procedure must be fitted.
Insufficient water hardness (below 8.4 °dH) can promote the growth of al-
gae.
Note
If necessary, insert a filter, water trap or air dryer.
Air and water requirements according to DIN EN 7494-2
The compressed air must be free of oil, dirt, and contamination. If needed:
▪ Use a compressor with a dry air system.
▪ Include an upstream air filter (on the compressor), if applicable.
▪ Blow out the lines before connecting them.

5.1.2 Connecting the ELECTROmatic

Property damage due to soiled water lines.
Property damage to the product.
▶ Make sure that the water filter does not get clogged by soiled water.
▶ For this purpose, rinse the turbine hose for at least 1 minute before con-
necting its free end to the product in order to remove any soiled water from
the lines.
NOTICE
NOTICE
NOTICE
48 / 80