5

Commissioning device

Mechanical installation and validation procedure
Procedure
1. Power up the rugged tablet PC and connect the CC21.
2. Double-click the Rosemount
installation app desktop icon. Within approximately 10 seconds, the
installation tool software should open.
3. Attach the CC21 to the sensor.
4. In the Rosemount WT210 in installation app software, as seen in
Figure
Note
Joining the device to the network could take several minutes.
Figure 5-1: Rosemount WT210 Install Tool
12
5-1:
a) Select the Provision tab.
b) Enter the five-digit network ID and the 32 hexadecimal
(numbers 0-9 and letters A-F) join key.
c) Click the Provision button. Confirmation is given when
provisioning is complete.
d) Check in the Network Discovery panel to confirm that the
sensor can hear a device with the network ID you wish the
sensor to join.