July 2019
Figure 4-3: Thermocouple Guidelines
6. Locate and place the WT210 sensor over the studs, ensuring the
waveguides are positioned centrally to +/-0.2 in. (5 mm).
7. Place washers over the studs.
8. Thread nut onto the studs and run them down several threads.
9. Carefully secure the sensor in this position by finger tightening each
nut after observing the sensor in
Figure 4-4: Equal Spacing Guidelines
Quick Start Guide
Correct
Incorrect
Figure 4-4.
Quick Start Guide
11