Section 9.0
Network Connections
Broad-band is the standard network connection and is required for BrightSpeed Series systems. (A
dial-up modem is optional.) A 1000 baseT high-speed network is desired, with 100 baseT network
service acceptable. Broad-band connections should use one of the following Category 5 patch
cables:
CAT Num
K9000WB
K9000KP
K9000JR
K9000WA
The CT system is connected to the network through the Console.
A patch cable (not to exceed 3.05 m (10 ft. ) should be provided by the customer, and it is used
to connect the console to a wall box. (See Notes on
Some customer-site units may require cable duct-work or conduit to route connecting network
cables to the workstation, camera and console.
The run from the hospital switch to the CT wall outlet must not exceed 88m (290 ft. ). Bandwidth
performance is degraded when the length reaches 91m (300 ft. ) or greater.
For the optional modem: Two phone lines should be provided by the facility. One line is for
use with a modem and must be an analog line. The second line is a voice only line.
Hospital Network
Figure 4-19 LCGOC/GOC4/AIO Console Rear Bulkhead
14
GE Part Num
2215028-10
2215028-5
2215028-4
2215028-9
Advantage Window
Chapter 4 - Room Planning
B
S
E
, E
RIGHT
PEED
LITE
Length
20 m
10 m
5 m
3 m
Figure
9-5)
, E
: P
-I
DGE
XCEL
RE
NSTALLATION
Page 87