GE BrightSpeed Elite Preinstallation Manual page 144

Table of Contents

Advertisement

LENGTH,
ACTUAL
(USABLE)
PART # DESCRIPTION
ft
m
050
63
19.3
2343529
HVDC, PDU to Gantry
(55)
(16.76)
051 62.5
19.3
2343530
HVAC, PDU to Gantry
(55)
(16.76)
052
63
19.3
2343528
LVAC, PDU to Gantry
(57)
(16.76)
053
80
24.5
2343531
LVAC, PDU to Console
(75)
(22.86)
055
63
19.3
2371450
Ground, PDU to Raceway
(55)
(16.76)
056
83
25.5
2371450-3 Ground, Raceway to Console 1283 VW-1
(75)
(22.86)
100
70
21.4
5120646
Signal, Gantry MSUB to PDU
(62)
(18.86)
101
86
26.35
5120645
Signal, Gantry MSUB to OC
(78)
(23.71)
102
86
26.3
2373436-2 Signal (LAN), Gantry to OC
(81)
(24.86)
103
80
24.3
2117848-2 Fiber Optic, Gantry to OC
(75)
(22.86)
Table 9-9 GE Healthcare Supplied Cables for GOC or TIO (Optional, Long Run) (2281840-4)- UL Information
UL CABLE INFORMATION
2587 FT4
600
+ & - 350VDC 90
2587 FT4
600
440Y/254
2587 FT4
600
208Y/120
2587 FT4
600
120VAC
1284 VW-1
600
0
(FT-1)
600
0
(FT-1)
FT-4
300
<30VDC
FT-4
300
<30VDC
1900 <30VDC
N/A
N/A
PULL SIZE
MM (INCHES)
19 (0.751) 3
(2) 4
22 (0.87) Dia
(1) 8
90
15.3 (.604) 4
14
11.2 (0.44) Dia
90
13.8 (.542) 5
8
56.4 (2.22) Dia
90
12.3 (.483) 3
10
56.4 (2.22) Dia
105
15.5 (.608) 1
1/0
15.8 (0.62) Dia
105
11.9 (.467) 1
2
12.2 (0.48) Dia
80
11.2 (.440) 25 22
17 x 58 (0.68 x 2.30)
19 x 51 (0.75 x 2.01)
80
11.2 (.440) 25 22
17 x 58 (0.68 x 2.30)
19 x 51 (0.75 x 2.01)
5.9 (.234)
8
24
15 (0.59) Dia
1
N/A
10 (0.39) Dia

Advertisement

Table of Contents
loading

This manual is also suitable for:

Brightspeed edgeBrightspeed excel

Table of Contents