Table 4-7 UPS Subsystem
Page 56
14
Work Space 
Requirement
Direction of Service
Access (Front of UPS)
Service Access Width
(Right side and length of
UPS)
Head Clearance
Work Space 
Requirement
Direction of Service
Access (Front of A1
Disconnect)
Service Access Width
(Right side and length
of A1 Disconnect)
Head Clearance
Table 4-8 A1 Disconnect Subsystem
Section 2.0 - Regulatory Clearances
B
RIGHT
Minimum Clear
Additional Conditions
Space
914.4 mm (36 in.)*
There are no exposed live part hazards with
the cover in place.
This component is typically serviced from
the front with access to the rear.
*If exposed live parts of 151 - 600 volts are
present, 1219 mm (48 in.) is required on
both sides of the workspace with the opera-
tor between.
*If the opposite wall is grounded and
exposed live parts of 151 - 600 volts are
present, 1067 mm (42 in.) is required.
762 mm (30 in.)
This is the width of the working space in
front of the equipment. A minimum of 762
mm (30 in.) or the width of the equipment,
whichever is greater, is required.
1981 mm (78 in.)
This is the height of the workspace
measured from the floor at the front edge of
the equipment to the ceiling or overhead
obstruction(s). A minimum of 1981 mm
(78 in.) or the height of the equipment,
whichever is greater, is required.
Minimum Clear
Additional Conditions
Space
914.4 mm (36 in.)*
There are no exposed live part hazards with
the cover in place.
This component is typically serviced from
the front with access to the rear.
*If exposed live parts of 151 - 600 volts are
present, 1219 mm (48 in.) is required on
both sides of the workspace with the opera-
tor between.
*If the opposite wall is grounded and
exposed live parts of 151 - 600 volts are
present, 1067 mm (42 in.) is required.
762 mm (30 in.)
This is the width of the working space in
front of the equipment. A minimum of 762
mm (30 in.) or the width of the equipment,
whichever is greater, is required.
1981 mm (78 in.)
This is the height of the workspace
measured from the floor at the front edge of
the equipment to the ceiling or overhead
obstruction(s). A minimum of 1981 mm
(78 in.) or the height of the equipment,
whichever is greater, is required.