Section 2.0

Regulatory Clearances

2.1
Regulatory Clearances
MINIMUM CLEARANCES UNDER U.S. FEDERAL REGULATIONS AND NATIONAL
STANDARDS: 29 CFR 1910 (OSHA), NFPA 70E (STANDARD FOR ELECTRICAL
SAFETY IN THE WORKPLACE), AND NFPA 101 (LIFE SAFETY CODE):
Figure 4-1
on the following pages for detailed dimensional clearances. Please note all systems installed in the
United States must comply with all Federal and local regulations. For installations outside the
United States, country-specific or other local regulatory clearance requirements must be met.
A) Minimum Room
B) Minimum Room
C) Recommended Room
Reference Regulatory Clearance Tables
for minimum and variable distance definitions
Figure 4-1 Regulatory Clearance Requirements for BrightSpeed System Configurations
Note:
See
Section 3.0
Page 52
14
is a map of clearance requirements for U.S. regulatory compliance. See clearance tables
NGPDU
914 to 1219mm
(36 to 48 in.)
Depth
762mm(30in.)
Width
1981mm(78in.)
Height
(Less than 711 mm (28 in. ))
(Less than 914 mm (36 in. ))
(914 mm (36 in.) or Greater)
Regulatory Compliance
Workspace Clearance
Perpendicular to Equipment
Regulatory Compliance 711 mm
28 inch Egress Clearance
on following pages
and additional metric conversions.
for Service Clearances
Section 2.0 - Regulatory Clearances
B
S
RIGHT
2057 mm
( 6ft. 9in. )
Cradle
Extender
C B
A
Gantry Super
Structure
914 to 1219mm
(36 to 48 in.)
Footprint
Foot
Switch
Asm
Table
Base
Table
Super
Structure
Footprint
610 mm
(24 in.)
E
, E
, E
: P
PEED
LITE
DGE
XCEL
RE
&
711 mm
(28 in.)
914 to 1219mm
(36 to 48 in.)
914 mm
(36 in.)
914 mm
(36 in.)
711 mm
(28 in.)
-I
NSTALLATION