GE BrightSpeed Elite Preinstallation Manual page 55

Table of Contents

Advertisement

GE H
EALTHCARE
D
5141177-100, R
IRECTION
EVISION
Work Space
Requirement
Service Access Width
(Left-Right of workspace)
Table 4-5 Gantry Subsystem
Work Space
Requirement
Direction of Service
Access (Table Head or
Foot)
Direction of Service
Access (Table Sides)
Direction of Service
Access (Table Foot)
Service Access Width
(Left-Right of workspace)
Table 4-6 Table Subsystem
14
Minimum Clear
Space
762 mm (30 in.)
Minimum Clear
Space
914 mm (36 in.)
914 mm (36 in.)*
711 mm (28 in.)
762 mm (30 in.)
Chapter 4 - Room Planning
B
S
E
, E
, E
RIGHT
PEED
LITE
DGE
XCEL
Additional Conditions
This is the width of the working space in
front of the equipment. A minimum of 762
mm (30 in.) or the width of the equipment,
whichever is greater, is required.
Additional Conditions
There are no exposed live parts hazards
with the cover in place.
This component is typically serviced form all
four sides.
this is the width of the workspace on each
side of the equipment. A minimum of 914.4
mm (36 in.), or the width of the equipment,
whichever is greater, is required.
*This distance can be reduced to 711 mm
(28 in.) provided a written and signed
approval is obtained by the local team from
the local AHJ (Authority Having
Jurisdiction). The signed document must be
on file with GE.
For the front gantry cover removal, a
minimum of 457 mm (18 in.) is allowed only
if an unobstructed egress space of 711 mm
(28 in.) is maintained around the equipment
for room exit. This also means no trip
hazards exist along the path of egress.
This is the width of the working space in
front of the equipment. A minimum of 762
mm (30 in.) or the width of the equipment,
whichever is greater, is required.
: P
-I
RE
NSTALLATION
Page 55