6.0 Maintenance and service

Maintenance, repairs and period tests may only be carried out by persons who have
the appropriate technical knowledge and are familiar with the product. The person in
question must possess the necessary test devices and genuine spare parts required to
carry out these measures.
ATMOS recommends: Work should be carried out by an authorised ATMOS service partner.
Period tests
Comply with the country-specific guidelines regarding regular testing, especially with
regard to electrical safety. ATMOS recommends performing a test every 24 months.

6.1 Function check

Perform a function check:
Prior to each use.
After each use and after cleaning.
Every 4 weeks if the device is not being used.
After any maintenance work, servicing or repair.

6.2 Sending in the device

Remove all consumables and dispose of them properly.
Clean and disinfect the product and accessories in accordance with the operating
instructions.
Place any used accessories with the product.
Fill in the QD 434 "Delivery complaint/return shipment" form and the corresponding
decontamination certificate.
This form is enclosed with each delivery and can be found at www.atmosmed.com.
The device must be well padded and packed in suitable packaging.
Place the QD 434 "Delivery complaint/return shipment" form and the corresponding
decontamination certificate in an envelope.
Affix the envelope to the outside of the package.
Send the product in to ATMOS or your dealer.

6.3 Handling batteries

The battery in the ATMOS LC 27 Battery is a wear part and is therefore not covered by
the general warranty. There is a function guarantee of 6 months.
Please observe the following notes in order to reach the maximum service life of your
battery:
Only use the original battery.
Prior to first use the battery must be fully charged.
Battery-run devices should only be stored when they are charged.
Fully charge the battery every 3 months, even if the device is not used.
Do not expose the batteries to direct solar radiation and keep them away from
radiators. The perfect storage temperature is between 8 and 15° C.
Exchange the battery when the remaining battery service life noticeably decreases.
Batteries are run-down after approx. 400 charging cycles.