Section 14: TSP command reference

bufferVar.formattedreadings

This attribute contains the stored readings shown as numbers with units and prefixes.
Type
TSP-Link accessible
Attribute (R)
Yes
Usage
reading = bufferVar.formattedreadings[N]
reading
bufferVar
N
Details
This read-only attribute is an array that contains the stored readings. The readings are shown as
numbers with prefixes before the units symbol.
Example
reset()
testData = buffer.make(50)
trigger.model.load("SimpleLoop", 3, 0, testData)
trigger.model.initiate()
waitcomplete()
print(testData.formattedreadings[1])
printbuffer(1, testData.n, testData.formattedreadings)
Create a reading buffer named testData, configure the instrument to make three measurements, and store
the readings in the buffer.
Print the first reading.
Example output:
+00.0028 nA
Print all readings in the reading buffer.
Example output:
+00.0028 nA, +00.0039 nA, +00.0040 nA
Also see
bufferVar.readings
buffer.delete()
buffer.make()
bufferVar.clear()
print()
(on page 14-99)
printbuffer()
(on page 14-100)
Reading buffers
Remote buffer operation
14-24
Affected by
Restore configuration
Instrument reset
Power cycle
Buffer reading formatted with numbers and prefixes with units for element N
The name of the reading buffer, which may be a default buffer (defbuffer1 or
defbuffer2) or a user-defined buffer
The reading number N; can be any value from 1 to the number of readings in the
buffer; use the bufferVar.n command to determine the number of readings in the
buffer
(on page 14-28)
(on page 14-9)
(on page 14-12)
(on page 14-18)
(on page 6-1)
(on page 6-25)
2470 High Voltage SourceMeter Instrument Reference Manual
Where saved
Not applicable
Default value
Not applicable