Section 14: TSP command reference

tsplink.line[N].reset()

This function resets some of the TSP-Link trigger attributes to their factory defaults.
Type
TSP-Link accessible
Function
Yes
Usage
tsplink.line[N].reset()
N
Details
The tsplink.line[N].reset() function resets the following attributes to their default values:
tsplink.line[N].mode
trigger.tsplinkin[N].edge
trigger.tsplinkout[N].logic
trigger.tsplinkout[N].pulsewidth
trigger.tsplinkout[N].stimulus
This also clears trigger.tsplinkin[N].overrun.
Example
tsplink.line[3].reset()
Also see
trigger.tsplinkin[N].edge
trigger.tsplinkin[N].overrun
trigger.tsplinkout[N].logic
trigger.tsplinkout[N].pulsewidth
trigger.tsplinkout[N].stimulus
tsplink.line[N].mode
14-272
Affected by
The trigger line (1 to 3)
(on page 14-262)
(on page 14-263)
(on page 14-265)
(on page 14-266)
(on page 14-267)
(on page 14-271)
2470 High Voltage SourceMeter Instrument Reference Manual
Where saved
Resets TSP-Link trigger line 3 attributes to
default values.
Default value
2470-901-01 Rev. A / May 2019