Section 12: SCPI command reference
Example
TRACe:STAT:AVERage?
TRACe:STAT:AVERage? "testData"
Also see
Reading buffers
Remote buffer operation
:TRACe:MAKE
:TRACe:STATistics:CLEar
:TRACe:STATistics:MAXimum?
:TRACe:STATistics:MINimum?
:TRACe:STATistics:PK2Pk?
:TRACe:STATistics:STDDev?
:TRACe:STATistics:CLEar
This command clears the statistical information associated with the specified buffer.
Type
Command only
Usage
:TRACe:STATistics:CLEar
:TRACe:STATistics:CLEar "<bufferName>"
<bufferName>
Details
This command clears the statistics without clearing the readings.
Example
TRACe:STATistics:CLEar
TRACe:STATistics:CLEar "testData"
Also see
Reading buffers
Remote buffer operation
:TRACe:MAKE
:TRACe:STATistics:AVERage?
:TRACe:STATistics:MAXimum?
:TRACe:STATistics:MINimum?
:TRACe:STATistics:PK2Pk?
:TRACe:STATistics:STDDev?
12-134
(on page 6-1)
(on page 6-25)
(on page 12-127)
(on page 12-134)
(on page 12-135)
(on page 12-135)
(on page 12-136)
(on page 12-137)
Affected by
Not applicable
The name of the reading buffer, which may be a default buffer (defbuffer1 or
defbuffer2) or a user-defined buffer; if no buffer is defined, clears the statistics
from defbuffer1
(on page 6-1)
(on page 6-25)
(on page 12-127)
(on page 12-133)
(on page 12-135)
(on page 12-135)
(on page 12-136)
(on page 12-137)
2470 High Voltage SourceMeter Instrument Reference Manual
Returns the average reading for the readings in the default
buffer defbuffer1.
Returns the average reading for the readings in the
user-defined buffer testData.
Where saved
Not applicable
Clear all statistics in defbuffer1.
Clears all statistics in a user-defined buffer
named testData.
Default value
Not applicable
2470-901-01 Rev. A / May 2019
Need help?
Do you have a question about the Keithley SourceMeter 2470 and is the answer not in the manual?
Questions and answers