2470 High Voltage SourceMeter Instrument Reference Manual
[:SENSe[1]]:CONFiguration:LIST:STORe
This command stores the active measure settings into the named configuration list.
Type
Command only
Usage
[:SENSe[1]]:CONFiguration:LIST:STORe "<name>"
[:SENSe[1]]:CONFiguration:LIST:STORe "<name>", <index>
<name>
<index>
Details
Use this command to store the active measure settings to a configuration index in a configuration list.
If the index parameter is not provided, the new settings are appended to the end of the list. The index
only stores the active settings for a single active measure function.
Configuration lists are not saved when the instrument is turned off or reset. To save a configuration
list, create a configuration script to save instrument settings, including any defined configuration lists.
Refer to
Instrument settings stored in a measure configuration list
measure settings that the instrument stores.
Example
:SENSe:CONF:LIST:STOR "MyConfigList"
:SENSe:CONF:LIST:STOR "MyConfigList", 5
Also see
Configuration lists
[:SENSe[1]]:CONFiguration:LIST:CREate
*SAV
(on page 12-9)
2470-901-01 Rev. A / May 2019
Affected by
Recall settings
Instrument reset
Power cycle
A string that represents the name of a measure configuration list
A number that defines a specific configuration index in the configuration list
(on page 4-83)
(on page 12-61)
Section 12: SCPI command reference
Where saved
Saved settings
(on page 4-86) for a complete list of
Stores the active settings of the instrument to
the end of the configuration list named
MyConfigList.
Stores the active settings of the instrument to
the configuration list named MyConfigList
in configuration index 5.
Default value
Not applicable