Section 12: SCPI command reference
Example
:SYST:EVEN:COUN? ERR
Also see
:SYSTem:CLEar
:SYSTem:EVENtlog:NEXT?
:SYSTem:EVENtlog:SAVE
:SYSTem:EVENtlog:NEXT?
This command returns the oldest unread event message from the event log.
Type
Query only
Usage
:SYSTem:EVENtlog:NEXT?
:SYSTem:EVENtlog:NEXT? <eventType>
:SYSTem:EVENtlog:NEXT? <eventType>, <eventType>
:SYSTem:EVENtlog:NEXT? <eventType>, <eventType>, <eventType>
<eventType>
Details
When an event occurs on the instrument, it is placed in the event log. The
:SYSTem:EVENtlog:NEXT? command retrieves an unread event from the event log. Once an event
is read, it can no longer be accessed remotely. However, it can be viewed on the front panel.
To read multiple commands, execute this command multiple times.
If there are no entries in the event log, the following is returned:
0,"No error;0;0 0"
If the event type is not defined, an event of any type is returned.
Note that if you have used :SYSTem:ERRor[:NEXT]? to check events,
:SYSTem:EVENtlog:NEXT? shows the next event item after the last error that was returned by
:SYSTem:ERRor[:NEXT]? You will not see warnings or information event log items that occurred
before you used :SYSTem:ERRor[:NEXT]?
If the event type is not defined, an event of any type is returned.
12-110
(on page 12-105)
(on page 12-110)
(on page 12-112)
Affected by
Not applicable
Limits the event log entries that are returned to specific types; set to:
Returns only the next error: ERRor
Returns only the next warning: WARNing
Returns only the next informational message: INFormational
Returns any event: ALL
2470 High Voltage SourceMeter Instrument Reference Manual
Displays the present number of errors in the instrument event
log.
If there are three errors in the event log, output is:
3
Where saved
Not applicable
Default value
Not applicable
2470-901-01 Rev. A / May 2019