Provide sufficient space around the analyzer to allow maintenance and servicing of this
unit.
2.2

Mounting

This procedure requires a minimum of two people to safely move and mount the analyzer.
Detailed system specifications
NOTICE
In accordance with IEC 60079-0: 2011 Clause 30.
glands for the three electrical conduits. Do not redesign or reposition these glands without
consulting Emerson.
WARNING
HEAVY ITEM
In accordance with IEC 60079-0: 2011 Clause 30.
Failure to properly handle the analyzer may cause injury to personnel.
Make sure the wall the analyzer is mounted on is solid, stable, and of a suitable material
to hold the weight of the analyzer.
Handle the analyzer with caution during unpacking, installing, maintaining, and
transporting to prevent crushing of hands, feet, or other body parts.
The analyzer weighs 176.4 lb. (80 kg).
Emerson recommends that a minimum of two people move and lift the analyzer.
Wear suitable protective gloves and protective footwear.
Make sure that the wall fixing points are capable of supporting a load of 176.4 lb. (80 kg)
each; this includes a x2 safety factor. All mounting points are 0.6 in. (15 mm) diameter
holes.
Procedure
1. Use safety approved equipment to lift the analyzer from the stable platform. One
2. Mount the analyzer using the factory fitted and pre-drilled wall mounting brackets
Rosemount CT5800
person guides the unit into position as the other person carefully operates the
lifting equipment.
(Figure
2-1) using four M8 x 1.25 (5/16 - 18 UNC) A2/A4 grade 70 bolts.
shows outline dimensional drawings of the analyzer.
Figure 2-4
September 2019
shows the positions of the
Install
37