Reference Manual
00809-0100-5800
2
Install
This section describes the correct installation procedure for the analyzer.
NOTICE
This section is in accordance with IEC 60079-0: 2011 Clause 30. This section must not be
changed, amended, or removed.
2.1

Site requirements

Select an appropriate site for installing the analyzer.
NOTICE
This section is in accordance with IEC 60079-0: 2011 Clause 30. This section must not be
changed, amended, or removed.
WARNING
INSTALLATION REQUIREMENTS
The place of installation must be clean, dry, and protected against strong vibrations
and frost.
Observe the advisable operating temperatures given in
Do not subject analyzer to direct sunlight or heat sources.
For outdoor installation, Emerson recommends installing the analyzer in a cabinet. At a
minimum, protect it against rainfall.
WARNING
FIRE AND EXPLOSION
In accordance with IEC 60079-0: 2011 Clause 30.
Failure to ensure the area is safe or leaving the device powered on can cause an explosion
or fire and seriously injure personnel.
Do not open the Ex d enclosure unless the area is known to be free of flammable materials
or unless all devices are switched off.
Rosemount CT5800
September 2019
Detailed system
specifications.
Install
35