21. If you wish to perform a span for another gas, press
22. Repeat steps
23. If you wish to zero verify, press
24. If you are finished verifying the analyzer, press
The span verification procedure is now complete.
Rosemount CT5800
Figure 5-14: Calibration / Verification Screen
The software returns to the Select gas screen
Step 8
through
Zero
verification.
The software returns to the Gas Sensor Main screen
(Figure
Step 16
for the next gas.
. Then follow the zero verification procedure in
.
(Figure
Verifying gas concentrations
September 2019
.
5-1).
4-5).
79