Troubleshooting and diagnostics
September 2019
18. Apply power to the analyzer.
7.17

Closing the enclosure

After all internal connections have been made and proved/tested, close the enclosure and
secure the 20 off M16 x 45 captive screws in accordance with the torque sequence shown
below.
Procedure
1. Remove all protective masking tape covering the flamepath and enclosure sealing
2. Inspect the seal. If any damage is detected, stop the process immediately and
3. Clean the seal to remove any adhesive from the protective cover.
4. Inspect the flanges/flamepath for damage, i.e., scratches, chipping, or other forms
5. Carefully lift up the housing and align with the rear housing.
6. Make sure that all holes are aligned.
160
gaskets.
report to maintenance.
of damage/deformation. If any damage is detected, stop the process immediately
and report to maintenance.
If necessary, use the two holes to assist alignment.
Figure 7-46: Alignment Holes
Reference Manual
00809-0100-5800
Emerson.com/Rosemount