ATEX / IECEx Assessment Report Summary
September 2019
Table A-4: North America
Detail
Serial number
Year of manufacture
Warnings
Maximum supply pressure
A.9
Conditions of certification/special conditions
for safe use
This equipment has flamepaths which differ from those in IEC60079-1/EN 60079-1.
Cascade Technologies Ltd. shall be contacted for guidance when maintaining the
flamepaths.
The fasteners which secure the cover are non-standard and shall therefore only be
replaced by fasteners supplied by the manufacturer for this purpose. The fasteners
must always be fitted with the washer supplied by the manufacturer.
The equipment has non-conductive surfaces which are a potential electrostatic
charging hazard - see instructions for guidance.
The user shall ensure that the flow process gas is limited to a maximum flow rate of 6
liters per minute.
The equipment shall only be uses with process gases which are classified for equipment
group IIB + H2 and must not contain oxygen or any other oxidizer in concentrations
greater than those found in normal air.
North American conditions
The equipment has flameproof joints with dimensions which are other than those
specified in Table 2 of ANSI/UL 60079-1: 6th edition and Table 3 of CSA C22.2
60079-1:16. These flameproof joints are not intended to be repaired but where
necessary the original manufacturer shall be contacted for guidance and information
on the dimensions of the flameproof joints.
The fasteners which secure the cover are non-standard and shall therefore only be
replaced by fasteners supplied by the manufacturer for this purpose. The fasteners
must always be fitted with the washer supplied by the manufacturer.
The equipment has non-conductive surfaces which are a potential electrostatic
charging hazard - see the instructions for guidance.
174
(continued)
CSA North American Marking
As applicable
As applicable
WARNING - DO NOT OPEN WHEN AN EXPLOSIVE ATMOSPHERE IS
PRESENT.
WARNING - POTENTIAL ELECTROSTATIC CHARGING HAZARD - SEE
INSTRUCTIONS.
WARNING – DO NOT OPEN WHEN ENERGIZED.
WARNING – A SEAL SHALL BE INSTALLED WITHIN 50 mm OF THE
ENCLOSURE.
1 Bar
Reference Manual
00809-0100-5800
Emerson.com/Rosemount