Receive All Data - Siemens SIWAREX WP321 Operating Instructions Manual

Electronic weighing system
Hide thumbs Also See for SIWAREX WP321:
Table of Contents

Advertisement

Figure 7-20
● Right-click on "Calibration parameter (DR3)" and select "Receive data record".
Figure 7-21
Checking the scale following calibration
If the scale is to be used for business purposes only, check it as follows:
1. The scale is unloaded and the display shows "0 kg".
2. Place one or more known test weights on the scale. To check other weight points of the
scale in addition to the calibration weight, do not use the calibration weight.
3. Check the displayed weight value in the SIWATOOL program.
4. Remove the test weights from the scale.
5. Check that the display of the scale is 0 kg again.
7.4.4

Receive all data

● Activate the "Receive all data" function in the "Communication" menu.
SIWAREX WP321
Operating Instructions, 08/2019, A5E33715669A-AD
SIWATOOL: Obsolete calibration parameters
SIWATOOL calibration parameters: Receive data record
7.4 Fast commissioning (Quick Start) with SIWATOOL
Commissioning
55