4.2.7

Test the loaded MWL

1. Keep the lifting arms raised, and add the other calibration weights
2. Check again to ensure there is 10 mm (0.38") minimum clearance [12.5 mm (0.5") on MSI
scales] between the side of the stacked weights and the vertical arms of the calibration
weight supports. If necessary, move the two crank bodies further from the calibration weight
supports to achieve this clearance.
3. Recheck the crank body alignment: if necessary, see Align the crank bodies to each other
(Page 19).
Milltronics MWL Weight Lifter
Operating Instructions, 11/2011, 7ML19985CR03-03
calibration weights applied
lifting arms raised
calibration weight support (2)
minimum clearance
- on MUS, MBS, MCS scales: 10 mm (0.38")
- on MSI scales: 12.5 (0.5")
vertical arm of calibration weight support
calibration weights
Installing/Mounting
4.2 Installation procedure
23