Installing/Mounting
4.2 Installation procedure
5. Ensure there is 10 mm (0.38") minimum clearance between the underside of the base-bar
weight and the top surface of the calibration weight support when the lifting arms are fully
raised. If necessary, shim the MWL to achieve this clearance.
①
②
③
④
6. Ensure there is sufficient clearance between the side of the base-bar weight and the vertical
arms of the calibration weight support arm, or support bracket: see dimensions shown below.
If necessary, move the two crank bodies further from the calibration weight supports to
achieve this clearance.
①
②
③
7. Recheck the crank body alignment: if necessary, see Align the crank bodies to each other
(Page 19).
22
10 mm (0.38") minimum clearance
top surface of calibration weight support
lifting arm raised
underside of base-bar weight
minimum clearance
- on MUS, MBS, MCS scales: 10 mm (0.38")
- on MSI scales: 12.5 mm (0.5")
vertical arm of calibration weight support
base-bar weight
Milltronics MWL Weight Lifter
Operating Instructions, 11/2011, 7ML19985CR03-03