1 General information
10
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
1.4.8

Important instructions after repair / maintenance

To ensure a safe operating condition for the device, the following
checks must be performed after repair / maintenance:
● Any connections that may have loosened during repair must be
reconnected. This applies in particular to lines for the power supply
and for the protective earth conductor. The protective earth
conductor resistor must be checked after attaching the covers.
– R < 100  m Ω between components connected to the protective
earth conductor and the protective earth conductor connection to
the power socket.
1.4.9

Safety labels

Fuses
NOTE
Use ONLY fuses of the same type!