Final Steps - Dentsply Sirona Primescan AC Service Manual

Hide thumbs Also See for Primescan AC:
Table of Contents

Advertisement

2 Service tools
22
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
2.2.7

Final steps

You must carry out the following work upon conclusion of the restore
processes in order that the device can be operated as previously:
1. Connect the device with the Internet to activate Windows. This will
occur automatically if the device has a permanent connection to the
Internet. In this case, the Windows updates must also be installed.
2. If the device is subsequently to be operated offline, establish a
short-term Internet connection; e.g. using a mobile telephone
system. Then open the Windows settings in the Start menu and
click on Update and Security, then Activation on the left. Activation
can be initiated here manually.
3. Install the required software (CEREC SW, inLab SW, etc.).
4. Check the interface parameters in the user software and, if
necessary, configure them (see relevant user manual).
Optional:
● Restoring previously backed-up data.
● Installation of updates and additional programs (e.g. Sidexis)
● Entering the necessary personal settings (passwords, etc.)
● Installation of the software for auxiliary devices (e.g. modem,
network, ISDN, etc.)