Calibrating The Sensor For Redox - ProMinent delta Operating Instructions Manual

Control module
Table of Contents

Advertisement

7.3

Calibrating the sensor for Redox

Main
Operation
proCONTROL
Set
Clear
Fig. 22: Calibrating the sensor for Redox
Measured variables for the control module delta®
Set
Sensor calibr.
Controller
Concentration
Aux.
Calibr. Redox
Buffer 1: 220 mV
Buffer 2: 465 mV
Calibr. Redox
Buffer 1: 200 mV
Buffer 2: 465 mV
Sensor calibr.
Calibr. Redox
Buffer 220 mV
Adjust no
Adjust yes
Offset: 0 mV
Meas.Value 225 mV
Buffer 465 mV
Adjust no
Adjust yes
Offset: 0 mV
Meas.Value 465 mV
A0323
41