Calibrating The Sensor For Chlorine; Preparing The Calibration Of The Sensor For Chlorine - ProMinent delta Operating Instructions Manual

Control module
Table of Contents

Advertisement

7.1

Calibrating the sensor for chlorine

Main
Operation
proControl
Set
Clear
Sensor calibr.
Adj.zero point.
DPD adjust
Calibr. values
Sensor calibr.
Adj.zero point.
DPD adjust
Calibr. values
Sensor calibr.
Adj.zero point.
DPD adjust
Calibr. values
Fig. 20: Calibrating the sensor for chlorine
7.1.1

Preparing the calibration of the sensor for chlorine

Measured variables for the control module delta®
Set
Sensor calibr.
Controller
Concentration
Aux.
Adj.zero point.
continue with P
Z. point [4.10] mA
DPD adjust
DPD: [3.00] ppm
Adjust. values
Z. point 4.10 mA
Slope 1.18 mA/ppm
CAUTION!
Correct sensor operation / Run-in time
Damage to the product or its surroundings
Correct measuring and metering is only possible if
the sensor is working perfectly
Please read the operating manual for the sensor
Please also read the operating manuals for the
fittings and other components used
It is imperative that the run-in times of the sensors
are adhered to
The run-in times should be allowed for when plan‐
ning initial operation
It may take a whole working day to run-in the
sensor
Necessity of calibrating the zero point
Calibration of the zero point is not generally necessary.
Calibration of the zero point is only necessary if the
sensor is operated at the lower limit of the measuring
range or if the 0.5 ppm sensor version is used.
DPD adjust
Z. point 4.10 mA
Slope 1.18 mA/ppm
Calibr. values
Z. point [4.10] mA
Slope 1.18 mA/ppm
A0321
37