Calibration Of Zero Point And Gradient - ProMinent delta Operating Instructions Manual

Control module
Table of Contents

Advertisement

Measured variables for the control module delta®
7.1.2

Calibration of Zero Point and Gradient

Calibrating the chlorine sensor: slope
38
During the calibration, the control module delta
„0" . The mA standard signal outputs are frozen. The
outputs to
reading frozen at the start of calibration is suggested as a DPD
value. The DPD value can be set using the arrow keys.
NOTICE!
Prerequisites for correct calibration of the sensor
gradient
The DPD method required by the metering medium
employed will be used
The run-in time for the sensor has been adhered to
There is permitted and constant flow at the flow
gauge
There is temperature equalisation between the
sensor and the sample water
There is a constant pH value in the permitted
range
The sensor is fitted, flushed with sample water and connected
electrically to the control module delta
There has to be adequate metering medium in the sample water
for calibration (> 2% of the measuring range of the sensor).
Remove sample water directly at the measuring point and using an
appropriate reference method (e.g. DPD, titration etc.), determine
the content of metering medium in the sample water in
Enter this value in the control module delta
1.
Select the calibration menu
the button
2.
Take a sample of water and perform a reference measure‐
ment immediately.
„DPD adjust" to be calibrated using the
3.
Select the unit
buttons
or
4.
The press the button
ð The current reading will now be frozen.
5.
When necessary, adapt the ppm value determined using
keys
and
ð The ppm value of the sensor shown in this display now
corresponds to the reading in
6.
The press the button
ð The display now shows the value determined for the zero
point and gradient. Refer to the Error Message table
should an error be displayed
Necessity of calibrating the zero point
Calibration of the zero point is not generally necessary.
Calibration of the zero point is only necessary if the
sensor is operated at the lower limit of the measuring
range or if the 0.5 ppm sensor version is used.
®
sets the actuating
and run-in.
®
„ppm" .
as follows:
®
[Sensor Calibration] . The press
„ppm" .