Measured variables for the control module delta®
7

Measured variables for the control module delta

Measured variable
Chlorine
The measuring range can be selected infinitely from 0.5 ... 20 ppm.
pH Measured variable
Measuring range
Display range
Reference temperature
Resolution
Redox measured variable
Measuring range
Resolution
36
WARNING!
Danger of incorrect metering
This can result in hazardous incorrect metering
During initial commissioning, the measured variable
and the measuring range of the sensor must be set
prior to calibration.
When working on or with sensors, also observe the
respective technical documentation of the sensors.
Measurement range of the sensors
You must adapt the measuring range of the controller
to the measuring range of the chlorine sensor used.
Preset measuring range (default)
10 ppm
Typical measuring range
0 ... 20 mA
At least pH -1.45 ... 15,45
+ 25 °C
0.01 pH
Typical measuring range
- 1,000 mV ... + 1,000 mV
1 mV
®