Air in Line – Alarm Test ( Channel 1)
1.
Insert an administration set in Channel 1 and close the pump door. Press the door until a click is
heard. Switch the pump on.
NOTE:
For instructions on how to load and prime an administration set refer to sections
"Loading the Administration Set" & "Priming the Line".
2.
The pump will display channel 1 and the program selected. In order to edit channel 2 press the
CHANNEL key once.
3.
The pump displays Channel 1 and the program selected. Press
setting. If air sensor is off, Press
4.
Set/change the infusion rate at 500ml/hr, using the numeric
keypad and press
the charger's display shows 500.
5.
Set Volume at 20 ml. press
6.
The pump display will show the accumulation screen where
all set data is shown, check the settings, if correct, press
to start operation
6.
Disconnect the IV line from the IV bag or turn the drip chamber upside down to allow an air
bubble greater than 5 mm to enter into the administration set.
7.
Air-in-line alarm will be activated as soon as the air bubble enters the set segment located
behind the pump door.
8.
Press
NOTE: Same tests should be performed for channel 2
Down Occlusion – Alarm Test (Channel 1)
1.
Insert an administration set in channel 1 and close the pump door. Press the door until a click is
heard.
2.
Make sure room temperature is 21-23
NOTE:
For instructions on how to load and prime an administration set refer to sections
"Loading the Administration Set" & "Priming the Line".
to confirm setting. Verify that
to turn the alarm off.
to confirm.
key.
o
c.
to show the last rate
Rate 500 ml/h
Press OK for Volume
Volume 20 ml
Press OK Key
Rate
500 ml/h
Volume left
20 ml
Infused
Press Ok to Start
.
0 ml