Air in Line – Alarm Test
1.
Insert an administration set and close the pump door. Press the door until a click is heard.
NOTE:
For instructions on how to load and prime an administration set refer to sections
"Loading the Administration Set" & "Priming the Line".
2.
Press
until the Self-Test Screen appears. If air sensor is off, press
Please contact your technician in order to enable the air sensor if disabled with no need.
3.
The pump displays the program selected and its status. The display will automatically change to
display the last rate setting.
Change the infusion mode to Continuous (See "Change
4.
Set up" Section for instructions). Set the infusion rate at
500ml/hr, using the Numeric keypad and press
confirm setting.
VET-PRO VIP 2000™ Color Vision Operator Manual
.
to exit
to
Continuous
10:00
Prime Volume
5
mL
Press
key to Start
Continuous
10:00
-
Priming
██████████████
1
Continuous
10:00
Pump Paused
Too Long
Press
key
Continuous
10:00
Rate
500
mL/hr
Adjust and press
4 – Operation
80%
███■
-
80%
███■
5mL
80%
███■
-
to confirm.
80%
███■
-
Key