Date - Time: the upper side of the card (identified by Date - Time) is used to update, if necessary,
the instrument date and time; selecting the "Update instrument Date-Time to PC Date-Time" item,
the instrument date and time are automatically updated to the PC date and time.
Sample time: it represents the interval between two subsequent loggings (sampling interval) and
can be chosen in the range between 1 second and 1 hour.
Start Mode: it indicates the mode logging will start: selecting the Manual item, logging will start
when the operator presses the LOG key (or LOGGING key, depending on models) on the instru-
ment. On the contrary, selecting the Start Mode=Date item, the Start Date-Time (delayed start) and
Stop Date-Time boxes (logging delayed stop) will become active: here you can enter logging start
and stop time and date.
Anyhow, in both cases, you can start and stop logging from PC.
Alarms: it manages the alarms of each channel. Selecting the different items, related alarms are en-
abled and you can enter in the dialog boxes the minimum and maximum values, or both of them.
Each channel can be enabled independently of the other.
39