Setting Time and Date of Automatic Logging Start.
See the sequence of figures shown at step P03.
Warning: access to this step of the program is denied if P02 step has been set on auto ("dat"):
to be able to edit time and date, P02 has to be set on "man".
After setting start and stop time and date, P02 has to be set on "dat" to enable the auto logging
mode.
Setting Time and Date of Automatic Logging stop.
Refer to the sequence of figures shown at step P03.
Warning: access to this step of the program is denied if P02 step has been set on auto ("dat"):
to be able to edit time and date, P02 has to be set on "man".
After setting start and stop time and date, P02 has to be set on "dat" to enable the auto logging
mode.
Setting Alarm Set-points for the First Input Channel (Temperature in HD206-1 and
HD206-2 models and Internal Temperature in HD207 and HD207-1 models).
When acquiring, if temperature exceeds set limits (either low or high), the
symbol lights up
in correspondence with the ALRM indication (on display): the indication remains even if tem-
perature then falls into the limits again. This function is active if enabled through the P07
program.
When at P06, press ENTER to set low temperature set-point ("LOU"): use T3 (
) and T4
(
) keys to set the desired value, corresponding to the alarm low limit, press ENTER (T1) to
confirm. Then set high temperature set-point ("HIG"): use T3 (
) and T4 (
) keys to set
the desired value, corresponding to the alarm high limit, press ENTER (T1) to confirm.
10