Delta HD207 Series Instruction Manual page 12

Minidatalogger
Table of Contents

Advertisement

P07
off
ENTER
ENTER
[ ]
[ ]
2
1
Note: the alarm indication remains even if, after going out of range, the measurement re-
turns to set limits. To reset previous alarm indications, set OFF and press ENTER (T1)
to confirm.
Setting Alarm Set-points for the Second Input Channel (relative humidity in HD206-1 and
HD206-2 models and External Temperature in HD207 and HD207-1 models).
When acquiring, if the measurement related to the second input channel exceeds set limits (ei-
ther low or high), the
symbol lights up in correspondence with the ALRM indication (on
display): the indication remains even if the measurement value then falls into the limits again.
This function is active if enabled through P09 program.
When at point P08, press ENTER to set the low set-point ("LOU"): use T3 (
) and T4 (
)
keys to set the desired value, corresponding to the alarm low limit, press ENTER (T1) to con-
firm. Then set the high set-point ("HIG"): use T3 (
) and T4 (
) keys to set the desired
value, corresponding to the alarm high limit, press ENTER (T1) to confirm.
See the sequence of figures shown at P06 step.
12