Operation DDS
!
Important
notes on safety for the
DDS canister system
Fig 7.
Fig 8.
Fig 9.
Fig 10.
● The device may not be operated
without DDS bacterial fi lter / over-
suction stop!
Therefore always hold ready a
replacement DDS bacterial fi lter!
● Before each use, check that the DDS
bacterial filter / oversuction stop is
clean and dry. Wet or dirty fi lters must
be replaced with new ones.
● Please use gloves when changing the
bacterial fi lter!
● The DDS bacterial fi lter / oversuction stop
is disposable.
An exchange of the DDS bacterial fi lter is
necessary before every change of patient.
In the case the device is exclusively used
in one patient, the fi lter must be exchanged
at least every two weeks(depending on
frequency of use).
● Vacuum connection
Direct-Docking-System
The vacuum connection between the
pump and the collection canister is
created automatically as soon as the
DDS canister is positioned correctly.
5.1
DDS canister and
DDS bacterial fi lter
With the DDS collection canister on a fi rm surface, position
the lid horizontally on top (the lid may not be twisted!)
Press down lightly onto the collection canister using both
hands until limit is reached (Fig 7).
5.2
Insertion/Removal of the
DDS canister
For removal, pull the DDS collection canister horizontally
outside; for insert it again, shift it horizontally onto the bac-
terial fi lter (Fig 8.).The fi lter must be fi xed to the housing
afterwards the secretion canister is inserted.
!
Please note that this order is strictly observed otherwise
power loss of the device could be the consequence!
If required the canister can be ejected even easier from
the device by means of a lever instrument (e.g. plain
spattle); see Fig. 9.
Tip
5.3

Connect hose

Press the required DDS hose adapter with 6 or 10 mm
diameter into the hole of the DDS collection canister lid
twisting slightly to ensure a tight fi t (Fig 10).
Twist slightly in the same manner when removing.