7.8 Logging

The appliance continually logs all relevant measured values, settings and error messages
at 1-minute intervals. The internal log memory is of the continuous memory type. The
logging function cannot be switched off and is always active. The measured data are
stored in the appliance, safe from manipulation. If the power supply is interrupted, the
time of the power failure and voltage recovery are stored in the appliance.
You can export the protocol data for different periods to a USB storage medium via the
USB port or, via Ethernet, import them to the AtmoCONTROL software for graphical rep-
resentation, print-out or storage.
The log memory of the appliance is not modified or deleted by reading it out.
1. Connect the USB storage medium to
the USB port on the right of the Con-
trolCOCKPIT.
2. Activate the protocol. To do so, press
the activation key on the right side of
PROTOCOL
the
enlarged and the period
automatically highlighted. To select
another logging period, use the turn
control.
3. Save your selection by pressing the
confirmation key. The transfer starts
and a status symbol indicates the
progress.
4. As soon as the transfer is complete,
a check mark appears in front of the
period selected. You can now remove
the USB storage medium.
For a description of how to import and process protocol data in AtmoCONTROL
or read it out via Ethernet, please refer to the separate AtmoCONTROL manual.
D39374 | Date 08/2018
display. The display is
This month
Protocol
control range
12.10.2012 12:00
This Month
This Year
Complete control range
Menu mode
This Month
This Year
Complete control range
61

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vo29Vo49Vo101

Table of Contents