Operation and control
2. Accept the selection by pressing the
confirmation key. The
dertemperature protection) is automati-
cally activated.
3. By turning the turn control, adjust the
desired lower alarm limit value, in the
example on the right 160 °C.
If no undertemperature protection limit
is required, set the lowest temperature.
4. Press the confirmation key to confirm.
max
The
display (overtemperature
protection) is activated.
5. By turning the turn control, adjust the
desired upper alarm limit value, in the
example on the right 190 °C.
The monitoring temperature must be
set sufficiently high above the maximum
set temperature. We recommend 5 to
10 K.
6. Accept the upper alarm limit value
by pressing the confirmation key. The
setting of the automatic temperature
monitor (ASF) is automatically activated
auto
(
).
7. With the turn control, select ON (
OFF ().
8. Press the confirmation key to confirm.
The ASF tolerance band setting is acti-
vated.
9. With the turn control, adjust the desired
tolerance band, e.g. 5.0 K.
We recommend a tolerance band of 5
to 10 K.
40
min
setting (un-
) or
ALARM
min
max
100
120
0 .
°C
.0 °C
auto
+ -
0.0
ALARM
min
max
160
120
0 .
°C
.0 °C
auto
+ -
0.0
ALARM
min
max
160
120
0 .
°C
0 .
auto
+ -
0.0
ALARM
min
max
160
190
0 .
°C
0 .
auto
+ -
0.0
ALARM
min
max
160
190
.0 °C
.0 °C
auto
+ -
0.0
ALARM
min
max
160
190
.0 °C
.0 °C
auto
+ -
0.0
ALARM
min
max
160
190
0 .
°C
.0 °C
auto
+ -
0.3
ALARM
min
max
160
190
0 .
°C
.0 °C
auto
+ -
5.0
D39374 | Date 08/2018
K
K
°C
K
°C
K
K
K
K
K

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vo29Vo49Vo101

Table of Contents