Memmert VO Series Operating Manual page 56

Vacuum oven
Hide thumbs Also See for VO Series:
Table of Contents

Advertisement

Menu mode
5. Set the calibration value to 0.0 K and
accept the setting by pressing the con-
firmation key.
6. Place the sensor of a calibrated refer-
ence meter on the thermoshelf to be
calibrated.
7. Close the door and, in manual mode,
adjust the set temperature to 120 °C.
8. Wait until the appliance reaches the set
temperature and displays 120 °C. The
reference instrument should display
122.6 °C.
9. In the SETUP, adjust the calibration
value Cal2 to +2.6 K (actual value
measured minus setpoint temperature)
and save the setting by pressing the
confirmation key.
10. After the calibration procedure, the
temperature measured by the reference
instrument should now also be 120 °C.
With Cal1, a calibration temperature below Cal2 can be programmed accordingly, and
with Cal3, a temperature above. The minimum clearance between the Cal values is 20 K.
If all calibration values are set to 0.0 K, the factory calibration settings are restored.
56
re
TEMP
100
12.09.2012
13:44
100,4 °C
4
Manueller Modus
.2
°C
3
100,1 °C
2
100,2 °C
1
Set 100.0
°C
TIMER
of °C
GRAPH
4
76
In 1
max
max
000°C
h
m
mb
min
mb
000°C
min
Set 50
16
+
auto off
Ende 13:30 23.11.
mb
-
99K
O
O
O N
O
O
O
ON
N
N
N
N
TEMP
120
.0°C
Set 120
.0 °C
re
TEMP
120
.0°C
Set 120
.0 °C
40.0
-
0,2
Cal1
C
120.0
0,0
Cal2
C
180.0
-
0,2
Cal3
C
TEMP
21
.4 °C
120
Set
.0 °C
122.6 °C
40.0
-
0,2
Cal1
C
120.0
+
2,6
Cal2
C
180.0
-
0,2
Cal3
C
120.0 °C
D39374 | Date 08/2018
K
K
K
K
K
K