Menu mode
7.3.5 Timer mode
Here, you can choose whether the digital backwards counter
with target time setting (see page 34, timer) should be
setpoint-dependent or not – this determines whether the
timer should not start until a tolerance band of ±3 K around
the set temperature is reached (Fig. 34, B) or whether it
should start immediately after activation (A).
A
Fig. 34
Timer mode
A
Timer independent of setpoint: Timer starts immediately after activation
B
Timer setpoint-dependent: Timer does not start until tolerance band is reached
If the temperature tolerance band is exceeded in setpoint-dependent mode, the timer
will be interrupted and only be resumed when the setpoint temperature is reached
again.
7.3.6 Remote control
In the setup entry remote control, you can set whether
the appliance should be controlled via remote control
and if so, in which mode. These settings are available:
Off
►
Read-only
►
Write
►
Write + Alarm
►
If the appliance is in remot control mode,
the symbol
appears in the temperature display. In
Write
the settings
cannot be controlled at the ControlCOCKPIT until the
remote control has been switched off (setting
Read only
set to
.
In order to use the remote control function, pro-
gramming skills and special libraries are required.
52
t
Timer
B
Write + Alarm
and
t
Timer
, the appliance
Off
) or
IP address
255. 145. 1 . 10
Subnet mask
255. 255. 0 . 0
Unit
°C
Inert gas
In 1
Timer mode
2/2
Remote Control
Off
Gateway
192.168.5 .1
TEMP
23
.2 °C
Set 38
D39374 | Date 08/2018
F
In 2
t
Setup
.0 °C