2. By turning the turn control to the left
or right, adjust the set value (e.g. to
180.0 ºC).
3. Save the set value by pressing the con-
firmation key.
The display returns to normal and the
appliance begins adjusting to the de-
fined set value.
The settings for further parameters and functions (pressure) can be made accordingly.
If no new values are entered or confirmed for approx. 30 seconds, the appliance
automatically restores the former values.
If you want to abort the setting procedure, press the
activation key on the left or right of the display that
you want to exit. The appliance restores the former
values. Only the settings that you have confirmed by
pressing the confirmation key before cancelling the
setting procedure are accepted.
5.3.3 Operating modes
The appliance can be operated in different modes:
Manual mode: The appliance runs in permanent operation at the values set on the
ControlCOCKPIT. Operation in this mode is described in chapter 5.3.4 .
Operation with digital backwards counter with target time setting, adjustable from 1
minute to 99 days (timer): The appliance will run at the values set until the set time
has elapsed. Operation in this mode is described in chapter .
program mode: The appliance automatically runs program sequences which have
been defined using AtmoCONTROL software at a computer / laptop and then trans-
ferred to the appliance from a USB stick or via Ethernet. Operation in this mode is
described in chapter 5.3.6 .
By remote control
The status display shows which operating mode or operating state the appliance is cur-
rently in. The current operating state is highlighted in colour and indicated by the text
display:
Appliance is in program mode
program is stopped
Appliance is in manual mode
The example on the right shows the appliance in
manual mode, identified by the coloured hand symbol.
D39374 | Date 08/2018
Operation and control
TEMP
22
Set
TEMP
23
Set 180
12.Sept.2012
13:44
Manual Mode
.4 °C
180
.0
°C
.2 °C
.0 °C
T
T
T
-
33

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vo29Vo49Vo101

Table of Contents