Menu mode
5. With the turn control, set the calibra-
Cal2
tion point
6. Confirm by pressing the confirmation
key. The corresponding calibration
value is automatically selected.
7. Set the compensation correction value
to 0.0 mb and confirm with the confir-
mation key.
8. In manual mode, set and activate set
pressure to 600 mb.
9. Wait until the set pressure is reached
and 600 mb is displayed. The refer-
ence instrument displays for example
607 mb.
SETUP
10. In the
, adjust the calibration
value Cal2 to +7 mb (actual value
measured minus setpoint value) and
save the setting by pressing the confir-
mation key.
11. After the calibration procedure, the
pressure measured by the reference in-
strument should now also be 600 mb.
58
to 600 mb.
Cal1 50.0 mb
Cal2
Cal2 950.0 mb
Cal1 50.0 mb
Cal2 600.0 mb
Cal2 950.0 mb
Cal1 50.0 mb
Cal2 600.0 mb
Cal2 950.0 mb
VACUUM
600
In 1
mb
Set 600
mb
40.0
Cal1 50.0 mb
Cal2 600.0 mb
Cal2 950.0 mb
VACUUM
600
In 1
mb
Set 600
mb
D39374 | Date 08/2018
40.0
0.0 mb
600.0
mb +4.0 mb
0.0 mb
40.0
0.0 mb
+4.0
mb
0.0 mb
40.0
0.0 mb
0.0
mb
0.0 mb
VACUUM
600
In 1
mb
Set 600
mb
607.0 mb
0.0 mb
+7.0
mb
0.0 mb
600.0 mb