Table 22.
Guidance and Manufacturer's Declaration —Electromagnetic Immunity
The BladderScan Prime system is intended for use in the electromagnetic environment specified below. The
customer or the user of the BladderScan Prime system should assure that it is used in such an environment.
IMMUNITY TESTS
Radiated RF
IEC 61000‑4‑3
Note: Ut is the AC mains voltage prior to application of the test level.
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio,
AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment
due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the
BladderScan Prime system is used exceeds the applicable RF compliance level above, the BladderScan Prime system should be observed
to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re‑orienting or relocating
the BladderScan Prime system.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
84
IEC 60601 TEST LEVEL
3 V/m
80 MHz to 2.5 GHz
COMPLIANCE
ELECTROMAGNETIC ENVIRONMENT –
LEVEL
d=1.2 √P 80 MHz to 800 MHz
d=2.3 √P 800 MHz to 2.5 GHz
where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
3 V/m
transmitters, as determined by an
electromagnetic site survey,
be less than the compliance level in
each frequency range.
Interference may occur in the vicinity
of equipment marked with the
following symbol:
GUIDANCE
should
a
b