DISINFECT THE PROBE
Low‑level disinfection of the probe dome (the gray area at the end of the probe) is required between uses.
Table 10.
Low‑Level Disinfection Methods for the Probe
PRODUCT
Metrex
CaviWipes™
®
17. Remove the gloves used in the cleaning portion of the procedure, and then put on new gloves.
18. Using a low‑level disinfectant from
according to the conditions in
probe dome as needed to ensure that it remains visibly wet for the duration of the exposure period. You
may use multiple wipes as necessary; do not reuse wipes.
Operations & Maintenance Manual: Cleaning & Disinfecting
Disinfect the
Probe. Low‑level disinfection of the probe dome is
LEVEL
Low‑level disinfect
Table
10, wipe the probe dome with the disinfection product
Table
10. The exposure period is 3 minutes. Using a new wipe, rewipe the
Probe dome
CONDITIONS
Exposure: Remain wet for 3 minutes.