12. Using cold, running water, rinse the lower portion of the probe for 1 minute. While rinsing, use the brush
to scrub the button, dome gasket, and label recess.
13. Using a new Metrex
30 seconds. Use a spiral pattern as indicated in the figure below. Repeat this process, using a new
wipe each time, to achieve the minimum cleaning time of 30 seconds.
14. Visually inspect the probe for contamination, paying particular attention to recessed or grooved areas. If
the probe is visually contaminated, repeat Step 9 though Step 13 until no contamination remains.
15. Allow the probe to air dry.
Note: If you are using a cart, place the probe into the cart with the dome facing upwards. This helps
prevent the dome from coming in contact with the cart.
52
CaviWipes,™ wipe the probe from the display toward the dome for at least
®
Start here with new wipe
Dispose of wipe