4. Using a screwdriver with a Phillips bit, unscrew the two screws securing the printer to the console. Stop
unscrewing when you encounter resistance; the screws are not designed to be completely removed
from the printer.
5. Lift the printer away from the console.
6. Using a cotton swab dipped in 70% isopropyl alcohol (IPA), clean the individual contacts on the printer.
Do not push the contact sideways or any farther than halfway down.
Note: Use only the minimum amount of IPA necessary to clean the contacts. Ensure that excess IPA
does not enter the interior of the printer.
7. Inspect the printer contacts and verify the following:
• The contacts are clean and free from corrosion, lint, or debris.
• The contacts appear to be straight.
• The contacts are uniformly spaced.