CLEAN THE PROBE CABLE & PROBE DISPLAY
6. Using a new Metrex
contaminants are removed.
CLEAN THE PROBE
7. Later in this procedure, you will use cold, running water. Position the console near the sink so the probe
can reach the water later when needed, but ensure the console is far enough away to stay dry.
8. Using a dry paper towel or soft cloth, wipe any ultrasound gel completely off the probe.
9. Using Metrex
wet, and hold for 2 minutes. Spray additional solution as needed to ensure that the probe remains wet
for the full 2 minutes.
10. Spray a soft‑bristled brush with Metrex
11. Using the brush, scrub the recesses around the probe button, dome gasket, and label for 30 seconds
each.
Operations & Maintenance Manual: Cleaning & Disinfecting
CaviWipes,™ wipe the entire probe cable and probe display until all visible
®
CaviCide,™ spray the lower portion of the probe (including button and label) until visibly
®
Probe button
CaviCide™ until the brush is thoroughly wet.
®
Probe dome gasket
Probe label
51