BLADDERSCAN BVI 3000 Verathon Service Partner Manual Effective: July 13, 2021 Caution: Federal (United States) law restricts this device to sale by or on the order of a physician.
Identifying the Source of the Issue ........................18 Procedure 4. Isolate the Service Issue ......................19 Repair Guide ..............................21 Troubleshooting ..............................24 Procedure 5. Identify Irregular Aiming Icons ..................... 24 Procedure 6. Test the Printer ........................25 Verathon Service Partner Manual: Table of Contents...
Page 6
REPAIR & REPLACEMENT ............................27 Minor Repairs ..............................27 Procedure 7. Clear a Paper Jam ........................ 27 Procedure 8. Clean the Printhead by Using Paper ..................28 Console Repair ..............................30 Procedure 9. Disassemble the Console ......................32 Procedure 10. Replace the EPROM ......................37 Procedure 11.
Page 7
Procedure 25. Run a Self Test ........................79 Procedure 26. Set Up & Maintain the String Phantom .................80 Procedure 27. Scan a Tissue‑Equivalent Phantom ..................82 Accuracy Specifications .............................85 Console Parts & Assemblies ..........................86 Part Numbers..............................87 Glossary ................................89 Verathon Service Partner Manual: Table of Contents...
BladderScan BVI 3000 devices. IMPORTANT For detailed information about the system and how to operate it, see the latest BladderScan BVI 3000 Operations & Maintenance Manual, available at verathon.com/product‑documentation. Prior to operating the instrument, ensure that you are familiar with the content in the Operations & Maintenance manual.
3.000 or higher. For software versions 3.000 and 3.010, it is recommended that you send the system to Verathon in order to have the EEPROM updated. Software versions under 3.020 have a maximum calendar date of December 31, 2019. If the software is not updated, the date on exam printouts may be inaccurate.
For any questions about RoHS‑compliant equipment, please contact Verathon Customer Care. UDI COMPLIANCE Depending on its date of manufacture, a Verathon product may or may not conform to the Unique Device Identification (UDI) standard. For items manufactured before the standard went into effect, the identity of the product line and the individual device are shown on the product’s labels and packaging as plain text.
INSTRUMENTS & TOOLS Be aware of the force that you apply to screws, as they or their housings can break if too much force is used. For information about the parts and components used when servicing a BVI 3000 system, see Part Numbers on page 87.
To avoid paper jams, never fold the end of the paper roll, cut it diagonally, or cut it to a point. WARNINGS WARNING Risk of explosion, fire, or serious injury. The BladderScan BVI 3000 system is battery‑powered. Failure to note the following when handling the battery may result in serious injury or damage: •...
ANNUAL CERTIFICATION When servicing an instrument, it is necessary to verify that the device is functioning properly prior to returning it to use. This chapter contains required certification procedures for the BVI 3000 system: • Enter Configuration Mode—Prior to performing repairs or certifying the instrument, it is necessary that you put the console in configuration mode.
• Test the Battery Charger & Batteries QUANTITY PART # DESCRIPTION Tools 0620‑0068 Certified adult bladder phantom 0400‑0132 BVI 3000 battery charger 0400‑0039 BVI 3000 battery 0800‑0042 Thermal paper roll for the printer 0620‑0156 BVI 3000 string phantom kit — Clean, soft cloth Verathon Service Partner Manual: Annual Certification...
Page 16
Location: The tests performed by the technician on the equipment listed above confirmed that it met all applicable product manufacturing and test specifications. Certification service is performed in compliance with Verathon Inc. manufacturing test specifications and equipment. Verathon recommends that certification of calibration be performed once a year.
Page 17
Note: The change in lighting may be subtle and difficult to see in a bright room. Light sensor 13. Remove your finger from the sensor. Confirm that the screen backlight turns off. Note: The change in lighting may be subtle and difficult to see in a bright room. Verathon Service Partner Manual: Annual Certification...
Page 18
EPROM ≥ 2.100 Optional Replace the EPROM EEPROM 3.000–3.010 Recommended Return the system to Verathon for update — EEPROM ≥ 3.020 Optional Return the system to Verathon for update — Software version has a maximum calendar date of December 31, 2019. If the software is not updated, the date on exam printouts may be inaccurate.
Page 19
24. Confirm that the test printout looks like the following simple grayscale pattern and alphanumeric characters. !"#$%&’() +,-./01234 56789:;<+>?@ABCDEFGHI JKLMNOPQRSTUVWXYZ[ ] _ʽabcdefghijklmnopqrs tuvwxyz{|}~üäåÄÅÉÆöõï Probe: 21308 Box: 00001965 VER: 3.020 D949 Probe Scan Count: 865 Console Scan Count: 1282 Verathon Service Partner Manual: Annual Certification...
Page 20
PERFORM A STRING TEST 25. Ensure that the string phantom has been properly set up according to the instructions in the procedure Set Up & Maintain the String Phantom on page 80. 26. Turn the console off, and then turn the console on. The console is now in configuration mode. 27.
Page 21
Table 1 on page 22. 34. Remove the probe from the calibration tank, and then dry it with a clean, soft cloth. 35. Restart the console. This returns the console to the Main Configuration screen. Verathon Service Partner Manual: Annual Certification...
Page 22
SETUP 38. Enter the desired next calibration date for the instrument, and then press the Set button. Note: Verathon recommends that you set the next calibration due date to 14 months from the date currently configured on the console. 12/2015 39.
Page 23
47. On the console screen, verify that the measurement is accurate to ± (10% + 10 mL) of the phantom’s volume. 48. On the certificate, make note of the scan result. 49. Repeat Step 42 through Step 48 in female mode, and then confirm the female mode measurement is within the accuracy range. Verathon Service Partner Manual: Annual Certification...
Page 24
50. Press the DONE button. 51. On the Scan Results screen, press the PRINT button. 21 cm 21 cm PRINT SCAN NOTES MAIN 52. On the Print screen, press the PRINT button. The console prints the scan results. LAST IMAGE COST SAVING TEST PRINT PRINT...
SEL+ SEL- EXIT SETUP 4. Restart the console, and ensure that the console starts in standard operating mode. Figure 4. Main Menu in Standard Operating Mode BVI 3000 VERATHON 05/22/14 13:27 SCAN PRINT NOTES Verathon Service Partner Manual: Annual Certification...
• Ensure that the user is following the proper instructions for use, as provided in the latest version of the Operations & Maintenance Manual, available at verathon.com/product‑documentation. • When scanning patients, ensure that the user is familiar with the following situations, which may affect accuracy: ○...
5. On the Main Configuration screen, press the RUN button. The console enters standard operating mode. THIS UNIT IS NOT SETUP CABLE PROG SETUP TEST PROBE TEST 6. On the Main Menu screen, press the ALT button. BVI 3000 VERATHON 05/22/14 13:27 SCAN PRINT NOTES Verathon Service Partner Manual: Diagnostics & Troubleshooting...
Page 28
17. Confirm the scan result is within an accuracy range of ± (10% + 10 mL). 18. If you suspect that the probe is leaking oil, contact Verathon Customer Care. 19. If a service issue occurs during this procedure, then service is required on the probe. Continue to the...
Please note that probe and console failures are not limited to those detailed in the table below. If you encounter an issue that is not documented in this manual or if you complete all of the corrective procedures and the service issue is not resolved, contact Verathon Customer Care. Table 1.
Page 30
Adjust the probe seating in the insert with a rocking or — twisting motion, and then rescan Probe String test fails † Contact Verathon Customer Care — Console Replace the PCB assembly in Reassemble the Console Complete the section Perform a String Test...
Page 31
If an error message occurs during testing and you are unable to reproduce it, continue service as normal. If the error message is reproducible, replace the PCB assembly. Observed failure typically occurs within the probe rather than the console. † Verathon Service Partner Manual: Diagnostics & Troubleshooting...
TROUBLESHOOTING PROCEdURE 5. IdENTIFY IRREGULAR AIMING ICONS The bladder shape displayed in the aiming icon on the LCD screen and on the printout should be shaped like an oval. Regardless of the volume measured, abnormalities in shape may indicate a service issue with the probe.
4. On the Print screen, press the SEL button until TEST PRINT is selected. LAST IMAGE COST SAVING TEST PRINT PRINT DONE 5. Press the PRINT button. The instrument begins printing. Note: Printing takes approximately one minute. LAST IMAGE COST SAVING TEST PRINT PRINT DONE Verathon Service Partner Manual: Diagnostics & Troubleshooting...
Page 34
6. If you would like to cancel printing while it is in progress, press the ABORT PRINT button. PRINTING IN PROGRESS ABORT PRINT 7. When printing is completed, tear the paper from the instrument by pulling the printout towards the back of the instrument.
7. If this procedure resolved the service issue and no additional service is required, continue directly to the Annual Certification chapter on page 6 in order to certify the system. If you were not able to clear the paper jam, complete the section Clear a Severe Paper Jam on page 42. Verathon Service Partner Manual: Repair & Replacement...
CABLE PROG SETUP TEST PROBE TEST 4. On the Main Menu screen, press the PRINT button. BVI 3000 VERATHON 05/22/14 13:27 SCAN PRINT NOTES 5. On the Print screen, press the SEL button until TEST PRINT is selected. LAST IMAGE...
Page 37
11. If this procedure resolved the service issue and no additional service is required, continue directly to the Annual Certification chapter on page 6 in order to certify the system. If this procedure did not resolve the service issue, see "Printout is irregular" in Table 1 on page 22. Verathon Service Partner Manual: Repair & Replacement...
CONSOLE REPAIR This section contains instructions for completing service on the BVI 3000 console. It consists of the following procedures: • Disassemble the Console (page 32) • Replace the EPROM (page 37) • Replace the LCD (page 38) • Replace the Window or Buttons (page 40) •...
PROCEdURE 9. dISASSEMBLE THE CONSOLE Use this procedure in order to disassemble the console to a stage from which you can complete any of the replacement procedures included in this section. If replacing a part or component requires additional disassembly, that information is detailed in the corresponding part‑replacement procedure. QUANTITY PART # DESCRIPTION...
Page 41
7. Carefully disconnect the three printer cables. Ensure that you do not damage, break, or bend the pins on the PCB. 8. If there is a ferrite suppressor around the printer cables, remove it. Verathon Service Partner Manual: Repair & Replacement...
Page 42
9. If you are completing any of the following procedures, Figure 15. (8) Screws in Analog PCB remove the (8) screws securing the analog PCB in place (Figure 15): • Service the Printer • Replace the Battery Contacts • Replace the Battery Eject Button •...
Page 43
If you are not completing any of the above procedures, skip directly to Step 15 of this procedure. 12. Remove the (8) screws securing the hide in place. 13. Flip the instrument over, and then open the paper door slightly. Verathon Service Partner Manual: Repair & Replacement...
Page 44
14. Hold down on the top cover, grab the paper door by the paper well, and then pull up. The hide and the paper door are released from the top cover. COMPLETE SERVICE PROCEdURES 15. Complete the service procedures in this chapter as needed, and then reassemble the console according to the procedure Reassemble the Console on page 54.
PCBs with a removable EPROM are software version 2.100 or lower. PCBs with a built‑in EEPROM are software version 3.000 or higher and would need to be sent to Verathon for any desired software updates. This procedure documents how to replace the EPROM on a PCB with a removable EPROM.
PROCEdURE 11. REPLACE THE LCd QUANTITY PART # DESCRIPTION Parts 0230‑0008 0200‑0147 1x2 socket strip 0200‑0139 2x10 socket strip 0264‑0003 LCD standoff 0260‑0264 LCD standoff nuts 0261‑0008 LCD standoff screws (#4‑40, ¼‑in, machine) Tools — Phillips bit — 3/16‑inch wrench or socket wrench REMOVE THE ORIGINAL LCd 1.
Page 47
7. Remove the protective film from the LCD. Ensure that you keep the LCD clean as you complete service and reassemble the console. 8. Complete any additional repairs to the console, and then reassemble the console according the procedure Reassemble the Console on page 54. Verathon Service Partner Manual: Repair & Replacement...
PROCEdURE 12. REPLACE THE WINdOW OR BUTTONS QUANTITY PART # DESCRIPTION Parts 0122‑0018 Window 0122‑0005 Buttons 0261‑0043 Window screw with washer (#4, ‑in, self‑tapping) Tools — Phillips bit 1. Complete the procedure Disassemble the Console on page 32. 2. Using a Phillips bit, remove the (7) screws securing the window to the top cover. 3.
Page 49
6. Using a Phillips bit and (7) window screws with washers (0261‑0043), secure the window to the top cover. 7. Complete any additional repairs to the console, and then reassemble the console according the procedure Reassemble the Console on page 54. Verathon Service Partner Manual: Repair & Replacement...
PROCEdURE 13. SERVICE THE PRINTER This procedure contains instructions for completing the following tasks: • Clearing a severe paper jam • Cleaning the print head and printer roller • Replacing the printer cable • Replacing the printer If the printer is producing irregular printouts, clean the print head and roller prior to replacing any parts. If cleaning the print head does not resolve the service issue, then replace the printer cable.
Page 51
9. If you are cleaning the printhead, use a cotton swab and isopropyl alcohol in order to wipe along the print head contact area. The contact area is the thin black line in the brown bar. Verathon Service Partner Manual: Repair & Replacement...
Page 52
10. If you are replacing the printer cable, carefully pull out the printer cable, and then place the printhead so that the printer cable port is facing directly up. Insert the new printer cable connector in to the port. Note: Do not press against the port on the printhead unless the port is supported from behind. This could cause the port to break off of the printhead.
Page 53
Black and blue wires 16. Seat the printer properly in the top cover. 17. Complete any additional repairs to the console, and then reassemble the console according the procedure Reassemble the Console on page 54. Verathon Service Partner Manual: Repair & Replacement...
PROCEdURE 14. REPLACE THE BATTERY CONTACTS QUANTITY PART # DESCRIPTION Parts 0263‑0001 Battery contact standoff nut (¼‑in, hex) 0130‑0075 Battery case contact 0261‑0007 Battery contact screw (#2‑56, ¼‑in, machine) Tools — Phillips bit REMOVE THE ORIGINAL BATTERY CONTACTS 1. Complete the procedure Disassemble the Console on page 32.
Page 55
9. Using (2) battery contact screws (0261‑0007) and a Phillips bit, secure the battery case contacts in place. 10. Complete any additional repairs to the console, and then reassemble the console according the procedure Reassemble the Console on page 54. Verathon Service Partner Manual: Repair & Replacement...
PROCEdURE 15. REPLACE THE BATTERY EJECT BUTTON QUANTITY PART # DESCRIPTION Parts 0122‑0009 Push rod for the battery eject button 0122‑0010 Push rod for the battery eject button 0261‑0030 Battery eject button screw (#4‑20, ‑in, self‑tapping) 0260‑0228 Washer Tools — File —...
Page 57
6. Mount the filed push rod on the top cover. Push rod 7. Place the rocker arm on top of the washer so that the knob of the push rod is located in the slot of the rocker arm. Rocker arm Verathon Service Partner Manual: Repair & Replacement...
Page 58
8. Using a Phillips bit, insert (1) battery eject button screw (0261‑0030) into the hole at the top of the rocker arm. Ensure that the screw is sufficiently loose to allow movement but tight enough to prevent rattling. 9. Push on the battery eject button and ensure that the mechanism is working as intended. 10.
3. On the top cover, use a file in order to reduce the plastic area that comes in contact with the printer door when the printer door is open. Note: Failure to sufficiently reduce the thickness of this plastic can cause the printer door to become stuck in the open position. Verathon Service Partner Manual: Repair & Replacement...
Page 60
ATTACH A NEW IR WINdOW 4. Remove the protective covering from both sides of the new IR window. 5. Slide the new IR window into the slot on the top cover. IR window 6. Apply a thin line of Loctite Bonder 422 along the two long, interior‑facing edges of the IR window. ATTACH AddITIONAL COMPONENTS When you replace the top cover of a BVI 3000 console, the battery contacts, battery eject button, printer, window, and buttons from the original console are frequently reusable.
6. Complete any additional repairs to the console, and then reassemble the console according the procedure Reassemble the Console on page 54. Verathon Service Partner Manual: Repair & Replacement...
PROCEdURE 18. REASSEMBLE THE CONSOLE Prior to reassembling the console, review the section Workmanship Standards on page 74. If any of the console components require replacement according to the guidance in that section, then replace those components during this reassembly procedure. QUANTITY PART # DESCRIPTION...
Page 63
8. Verify that the LCD is clean and free of dust, lint, or fingerprints. 9. Bend the ribbon cable between the PCBs in the assembly as shown in the following images. Figure 18. Curve for brown ribbon cable Figure 19. Curve for white ribbon cable Verathon Service Partner Manual: Repair & Replacement...
Page 64
10. Hold the PCB assembly at an angle, slide the Hirose connector into the gap in the top cover, and then set the PCB assembly in the top cover. 11. Ensure that the ribbon cable is not pressed against any screw bosses or other components. The pressure of being pushed against an object may damage the cable.
Page 65
If the digital PCB has an EPROM, insert (7) short PCB screws (0261‑0008) in the digital PCB as shown in the following figure. If the digital PCB does not have an EPROM, insert (4) short PCB screws (0261‑0008) into the digital PCB as shown in the following figure. Verathon Service Partner Manual: Repair & Replacement...
Page 66
16. If the instrument originally had a ferrite suppressor, place the ferrite suppressor around all of the printer wires and then clip it shut. Ensure that none of the wires are pinched between the two halves. 17. Connect the printer cables to the digital PCB. 18.
Page 67
19. If you did not remove the overmold, skip to Step 22. 20. Insert (4) large overmold screws (0261‑0031) near the back of the instrument. 21. Insert (6) small overmold screws (0261‑0030) near the front of the instrument. Verathon Service Partner Manual: Repair & Replacement...
Page 68
ATTACH THE BOTTOM COVER 22. Place the bottom cover on the overmold. 23. Using a Phillips bit, insert the (4) machine bottom cover screws (0261‑0013) that secure the bottom cover in place. 24. Ensure there are no gaps between the bottom cover, top cover, and handle. If needed, apply one or two bar clamps to the sides of the console, and then confirm the gap is closed.
Page 69
• If you replaced the EPROM or PCB assembly, continue to Program the Console on page 65. • If you do not need to replace the console label or program the console, continue to the chapter Annual Certification on page 6. Verathon Service Partner Manual: Repair & Replacement...
Fax: +31 (0) 20 210 30 92 0400-0039 Made in USA MADE IN USA To locate your nearest distributor, visit us online at www.verathon.com. US Patent No. 4926871, 5235985 INTL Patents Pending Medical - Ultrasound Equipment : ES60601-1:2005 and IEC 60601-2-37:2007 0125-2530-00-40 0125-0270...
Page 71
0123 Made in USA To locate your nearest distributor, visit us online at www.verathon.com. US Patent No. 4926871, 5235985 INTL Patents Pending 0125-0270 7. Peel back the protective backing on the console label in order to reveal the serial number window.
Page 72
9. Remove the protective backing from the console label, and then carefully apply the console label to the bottom of the console base housing, ensuring that the orientation is correct, as shown in the following figure. 10. If you replaced the EPROM or PCB assembly, continue to Program the Console on page 65.
S4 button beeps five times. DIGITAL TEST SELF TEST 6. Press the SELF TEST button. The console completes a self test, and then you are returned to the Manufacturing Test Menu screen. DIGITAL TEST SELF TEST Verathon Service Partner Manual: Repair & Replacement...
Page 74
7. On the Manufacturing Test Menu screen, press the MAIN button. MANUFACTURING TEST MENU AMODE BENCH SCAN TEST TEST MAIN 8. On the Main Configuration screen, press the SETUP NOW button. THIS UNIT IS NOT SETUP CABLE PROG SETUP TEST PROBE TEST 9.
Page 75
15. On the Date Preset screen, press the button to toggle between increasing or decreasing the month number, and then press the MM button until the correct month is displayed. Note: Only values from 01 to 12 are allowed. 05/22/14 DONE Verathon Service Partner Manual: Repair & Replacement...
Page 76
16. Press the button to toggle between increasing or decreasing the day, and then press the DD button until the correct day is displayed. Note: Only values from 01 to 31 are allowed. 05/22/14 DONE 17. Press the button to toggle between increasing or decreasing the year, and then press the YY button until the correct year is displayed.
After replacing the tailcone, continue to the Annual Certification on page 6. Figure 24. Replaceable Components of the BVI 3000 Probe Switch Overlay (0510‑0020) (Included in Tailcone Kit) Switch actuator (0120‑0658) Tailcone Kit (0800‑0479) (Included in Tailcone Kit) Verathon Service Partner Manual: Repair & Replacement...
PROCEdURE 21. REPLACE THE PROBE TAILCONE QUANTITY PART # DESCRIPTION Parts 0800‑0479 Tailcone Kit 0261‑0028 Tailcone screw (#8‑32, 1‑in, machine) 0510‑0020 Switch overlay 0120‑0658 0120‑0658 Switch actuator Tools — T15 star bit Included in tailcone kit. REMOVE THE TAILCONE 1. Using a T15 star bit, remove the two screws from the tailcone. 2.
Page 79
7. Peel the backing off of a switch overlay, orient it so that the waves are pointing towards the dome, and then apply the switch overlay onto the scan button. Switch overlay 8. Continue to the procedure Program the Tailcone on page 72. Verathon Service Partner Manual: Repair & Replacement...
PROCEdURE 22. PROGRAM THE TAILCONE QUANTITY PART # DESCRIPTION Tools 0570‑0090 BVI 3000 test console 0400‑0039 BVI 3000 battery Part Numbers on page 87 for orderable part number 1. Connect the fully assembled probe to a console that is operating in configuration mode. If needed, use a test console and the procedure Enter Configuration Mode on page 6 in order to put the system in...
A Probe Type = 12DOME_120 Phi Offset = 5 ID: 01234 Phi HYST = 2 Gain Offset = 0 # SCANS Phi Time = 2.000 – SAVE 5. Continue to the chapter Annual Certification on page 6. Verathon Service Partner Manual: Repair & Replacement...
REFERENCE DOCUMENTATION WORKMANSHIP STANDARDS When assessing whether a part needs to be replaced for cosmetic reasons, refer to the guidance in this chapter. PROBE COMPONENTS & PARTS Refer to the following guidance when assessing whether to replace a probe component: •...
Figure 30. Unacceptable Top Cover Cracking • Replace the applicable part if the bosses are broken or stripped. A boss is a channel or void in an extrusion that accepts a certain diameter screw. Verathon Service Partner Manual: Reference Documentation...
Page 84
• Replace the infrared (IR) window if it is cracked, damaged, or missing. If the IR window cannot be removed, the top cover must be replaced. • Replace the window if: ○ There are 3 or more significant cracks. ○ The window is significantly peeling. ○...
1. On the top of the control unit, open the paper well door. 2. If there is an empty paper roll in the paper well, remove it. 3. If there is a paper roll in the printer that is not Verathon part 0800‑0042, remove it. Verathon Service Partner Manual: Reference Documentation...
Page 86
4. If you are loading a new paper roll, cut off the first inch of the new paper. 5. Ensure that the end of the paper roll is cut straight. Do not fold the end of the paper roll, cut it diagonally, or cut it to a point.
As the test is completed, the Self Test screen displays the status of the systems tested. When the test is complete, the instrument returns to the Main Menu screen. ROM Test PASSED Bus Test PASSED NVRAM Test PASSED SRAM Test PASSED FLASH Test PASSED Verathon Service Partner Manual: Reference Documentation...
PROCEdURE 26. SET UP & MAINTAIN THE STRING PHANTOM The string phantom is used for aligning probes and testing proper probe function. The software in the probe detects the distance between the strings in the phantom. QUANTITY PART # DESCRIPTION Tools 0620‑0156 BVI 3000 string phantom kit...
Page 89
• Inspect the strings on a regular basis for any debris or contamination caused by impurities in the water. As needed, replace the water in the calibration tank. • In order to function properly, the phantom must be restrung when strings break or show signs of wear. Verathon Service Partner Manual: Reference Documentation...
• When not being used regularly, store the phantom in its case. • It is required that the tissue‑equivalent phantom be certified every year. If requested, Verathon will provide a loaner phantom free of charge while your phantom is being certified.
Page 91
At these high volumes, measurements are clinically useful even if they underestimate the true bladder volume. Verathon Service Partner Manual: Reference Documentation...
Page 92
12. If desired, repeat Step 7 through Step 9 two more times in order to ensure measurement consistency. 13. When you have collected sufficient scans in order to confirm aiming accuracy, press the DONE button. 132ml READY TO SCAN CURRENT = 132ml DONE SCAN CONFIRM MEASUREMENT ACCURACY...
The following accuracy range calculations are used when scanning human anatomy with a BVI 3000 instrument. Table 3. Scanning Accuracy Range BLADDER VOLUME ACCURACY RANGE 0 to 699 mL ± (20% + 20 mL) 700 to 999 mL ± (25% + 25 mL) Verathon Service Partner Manual: Reference Documentation...
Page 96
BVI 3000 Test System (order 0270‑0249) BVI 3000 test probe (order 0270‑0201) 0570‑0091 BVI 3000 battery charger 0400‑0132 BVI 3000 battery 0400‑0039 Thermal paper roll for printer 0800‑0042 Reference BVI 3000 Operations & Maintenance Manual (for the current edition, visit 0900‑0383 verathon.com/product‑documentation) Included in tailcone kit.
Liquid‑crystal display Printed circuit board Return authorization RoHS Restriction of Hazardous Substances Unique Device Identification, a service of the U.S. Food and Drug Administration (FDA) that tracks individual medical devices using standardized identification codes Verathon Service Partner Manual: Reference Documentation...